절연저항계

Analog Insulaton Testers 2406E

by 에스피티 2009. 11. 23.

Analog Insulaton Testers 2406E



The 2406E Analog Insulation Tester:

  • Analog models with two and three rating
  • AC voltage measurement
  • Automatic discharge
  • Sky blue EL backlight
  • Increased safety (covered battery charger)

Specifications - Model 240631

Rating125V/20M~
25V/5M~
50V/10M~
Effective Measuring Range0.001-5M~
0.005-10M~
0.01-20M~
Central Scale Value0.1M~
0.2M~
0.5M~
AC Voltage Measurement0 to 300 V
Lower Limit0.025M~
0.05M~
0.125~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Conditions - Model 240631

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.


Specifications - Model 240641

Rating125V/20M~
25V/5M~
50V/10M~
Effective Measuring Range0.001-5M~
0.005-10M~
0.01-20M~
Central Scale Value0.1M~
0.2M~
0.5M~
AC Voltage Measurement0 to 300 V
Lower Limit0.025M~
0.05M~
0.125~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Conditions - Model 240641

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.


Specifications - Model 240632

Rating125V/20M~
Effective Measuring Range0.01-20M~
Central Scale Value0.5M~
AC Voltage Measurement 0 to 300 V
Lower Limit0.125~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Condtions - Model 240632

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.


Specifications - Model 240642

Rating250V/50M~
Effective Measuring Range0.01-50M~
Central Scale Value1M~
AC Voltage Measurement 0 to 300 V
Lower Limit0.25~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Conditions - Model 240642

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.


Specifications - Model 240633

Rating125V/20M~
250V/50M~
500V/100M~
Effective Measuring Range0.01-20M~
0.01-50M~
0.05-100M~
Central Scale Value0.5M~
1M~
2M~
AC Voltage Measurement 0 to 600 V
Lower Limit0.125M~
0.25M~
0.5~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Conditions - Model 240633

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.


Specifications - Model 240343

Rating125V/20M~
250V/50M~
500V/100M~
Effective Measuring Range0.01-20M~
0.01-50M~
0.05-100M~
Central Scale Value0.5M~
1M~
2M~
AC Voltage Measurement 0 to 600 V
Lower Limit0.125M~
0.25M~
0.5~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Conditions - Model 240643

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.


Specifications - Model 240634

Rating1000V/2000M~
250V/50M~
500V/100M~
Effective Measuring Range0.01-50M~
0.05-100M~
1-2000M~
Central Scale Value1M~
2M~
50M~
AC Voltage Measurement 0 to 600 V
Lower Limit0.25M~
0.5M~
1~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Condtions - Model 240634

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.


Specifications - 240644

Rating1000V/2000M~
250V/50M~
500V/100M~
Effective Measuring Range0.01-50M~
0.05-100M~
1-2000M~
Central Scale Value1M~
2M~
50M~
AC Voltage Measurement 0 to 600 V
Lower Limit0.25M~
0.5M~
1~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Conditions - Model 240644

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.


Specifications - Model 240635

Rating1000V/2000M~
250V/500M~
500V/1000M~
Effective Measuring Range0.1-500M~
0.5-1000M~
1-2000M~
Central Scale Value10M~
20M~
50M~
AC Voltage Measurement 0 to 600 V
Lower Limit0.25M~
0.5M~
1~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Conditions - Model 240635

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading
Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.


Specifications - Model 240645

Rating1000V/2000M~
250V/500M~
500V/1000M~
Effective Measuring Range0.1-500M~
0.5-1000M~
1-2000M~
Central Scale Value10M~
20M~
50M~
AC Voltage Measurement 0 to 600 V
Lower Limit0.25M~
0.5M~
1~
Rated Current1 mA
Approx. Outer Scale Length88 mm
Discharge FunctionThe tester automatically begins discharge when you turn off the MEAS switch. The pointer swings if there is any residual voltage in the circuit under test. You can make sure discharge is complete by checking that the pointer swings back to the infinite (&
Approx. Width120 mm
Approx. Height110 mm
Approx. Depth60 mm
Approx. Weight500 g
BatteriesSix AA (R6P) batteries
Standard AccessoriesBatteries (Qty- 6)
Carrying case (B9075MU) (Qty -1) Earth probe (blake); approx. 1 m long
Earth probes and Line Probes (98007) (Qty - 1)
User's Manual (Qty - 1)- With probe-housing pocket and neck strap


Standard Test Conditions - Model 240645

Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
Position of UseHorizontal (5º max. of angle of inclination)
External Magnetic FieldsNone Battery voltage: Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length
AC voltage: ±10% of maximum scale value
No-load voltage: 130% max. of rated voltage
Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
Short-circuit current: 12 mA max.