본문 바로가기
절연저항계

Analog Insulation Testers MY10

by 에스피티 2009. 11. 23.

Analog Insulation Testers MY10 Series

 



Analog models with single rating
MY10-01:125V/20MΩ
MY10-02:250V/50MΩ
MY10-03:500V/100MΩ
MY10-04:500V/1000MΩ
MY10-05:1000V/2000MΩ

  • AC voltage measurement
  • Automatic discharge
  • A wide choice of accessories-Designed for shared use with the MY40


  • Specifications - Model MY10-01

    Rating125V/20MΩ
    Effective Measuring Range0.01-20MΩ
    Central Scale Value0.5MΩ
    AC Voltage Measurement 0 to 250 V
    Lower Limit0.125MΩ
    Rated Current1-1.2 mA
    Approx. Overall Scale Length107 mm
    Battery LifeApprox. 10 hours when continuously operated on manganese - oxide batteries with the pointer pointing to the central scale value.
    BatteriesFour AA (R6P) batteries
    Approx. Width125 mm
    Approx. Height103 mm
    Approx. Depth53 mm
    Approx. Weight500 g
    ComplianceEN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)
    Standard AccessoriesBatteries (Qty - 4)
    Earth Probe (98002) (Qty - 1)
    Line Probe (98001) (Qty - 1)
    Protection Cover (93013) (Qty - 1)
    Shoulder Strap (99005) (Qty - 1)
    User's Manual (Qty - 1)


    Standard Test Conditions - Model MY10-01

    Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
    Position of UseHorizontal (5° max. of angle of inclination)
    Battery VoltageWithin effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
    Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
    Second effective measuring range = ±10% of reading
    Infinite and zero scale values: 0.7% max. of scale length
    AC voltage: ±10% of maximum scale value
    No-load voltage: 130% max. of rated voltage
    Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
    Short-circuit current: 12 mA max.


    Additional Functions - Model MY10-01

    • Automatic discharge function
      • If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
      • When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.
    • Battery check (BAT mark on the scale plate)

     

    Specifications - Model MY10-02

    Rating250V/50MΩ
    Effective Measuring Range0.01-50MΩ
    Central Scale Value1MΩ
    AC Voltage Measurement 0 to 300 V
    Lower Limit0.25MΩ
    Rated Current1-1.2 mA
    Approx. Overall Scale Length107 mm
    Battery LifeApprox. 10 hours when continuously operated on manganese - oxide batteries with the pointer pointing to the central scale value.
    BatteriesFour AA (R6P) batteries
    Approx. Width125 mm
    Approx. Height103 mm
    Approx. Depth53 mm
    Approx. Weight500 g
    ComplianceEN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)
    Standard AccessoriesBatteries (Qty - 4)
    Earth Probe (98002) (Qty - 1)
    Line Probe (98001) (Qty - 1)
    Protection Cover (93013) (Qty - 1)
    Shoulder Strap (99005) (Qty - 1)
    User's Manual (Qty - 1)


    Standard Test Conditions - Model MY10-02

    Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
    Position of UseHorizontal (5° max. of angle of inclination)
    Battery VoltageWithin effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
    Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
    Second effective measuring range = ±10% of reading
    Infinite and zero scale values: 0.7% max. of scale length
    AC voltage: ±10% of maximum scale value
    No-load voltage: 130% max. of rated voltage
    Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
    Short-circuit current: 12 mA max.


    Additional Functions - Model MY10-02

    • Automatic discharge function
      • If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
      • When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.
    • Battery check (BAT mark on the scale plate)

     

    Specifications - Model MY10-03

    Rating500V/100MΩ
    Effective Measuring Range0.05-100MΩ
    Central Scale Value2MΩ
    AC Voltage Measurement 0 to 500 V
    Lower Limit0.5MΩ
    Rated Current1-1.2 mA
    Approx. Overall Scale Length107 mm
    Battery LifeApprox. 10 hours when continuously operated on manganese - oxide batteries with the pointer pointing to the central scale value.
    BatteriesFour AA (R6P) batteries
    Approx. Width125 mm
    Approx. Height103 mm
    Approx. Depth53 mm
    Approx. Weight500 g
    ComplianceEN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)
    Standard AccessoriesBatteries (Qty - 4)
    Earth Probe (98002) (Qty - 1)
    Line Probe (98001) (Qty - 1)
    Protection Cover (93013) (Qty - 1)
    Shoulder Strap (99005) (Qty - 1)
    User's Manual (Qty - 1)


    Standard Test Conditions - Model MY10-03

    Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
    Position of UseHorizontal (5° max. of angle of inclination)
    Battery VoltageWithin effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
    Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
    Second effective measuring range = ±10% of reading
    Infinite and zero scale values: 0.7% max. of scale length
    AC voltage: ±10% of maximum scale value
    No-load voltage: 130% max. of rated voltage
    Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
    Short-circuit current: 12 mA max.


    Additional Functions - Model MY10-03

    • Automatic discharge function
      • If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
      • When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.
    • Battery check (BAT mark on the scale plate)

     

    Specifications - Model MY10-04

    Rating500V/1000MΩ
    Effective Measuring Range0.5-1000MΩ
    Central Scale Value20MΩ
    AC Voltage Measurement 0 to 500 V
    Lower Limit1MΩ
    Rated Current0.5-0.6 mA
    Approx. Overall Scale Length107 mm
    Battery LifeApprox. 10 hours when continuously operated on manganese - oxide batteries with the pointer pointing to the central scale value.
    BatteriesFour AA (R6P) batteries
    Approx. Width125 mm
    Approx. Height103 mm
    Approx. Depth53 mm
    Approx. Weight500 g
    ComplianceEN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)
    Standard AccessoriesBatteries (Qty - 4)
    Earth Probe (98002) (Qty - 1)
    Line Probe (98001) (Qty - 1)
    Protection Cover (93013) (Qty - 1)
    Shoulder Strap (99005) (Qty - 1)
    User's Manual (Qty - 1)


    Standard Test Conditions - Model MY10-04

    Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
    Position of UseHorizontal (5° max. of angle of inclination)
    Battery VoltageWithin effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
    Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
    Second effective measuring range = ±10% of reading
    Infinite and zero scale values: 0.7% max. of scale length
    AC voltage: ±10% of maximum scale value
    No-load voltage: 130% max. of rated voltage
    Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
    Short-circuit current: 12 mA max.


    Additional Functions - Model MY10-04

    • Automatic discharge function
      • If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
      • When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.
    • Battery check (BAT mark on the scale plate)

     

    Specifications - Model MY10-05

    Rating1000V/2000MΩ
    Effective Measuring Range1-2000MΩ
    Central Scale Value50MΩ
    AC Voltage Measurement 0 to 500 V
    Lower Limit2MΩ
    Rated Current0.5-0.6 mA
    Approx. Overall Scale Length107 mm
    Battery LifeApprox. 10 hours when continuously operated on manganese - oxide batteries with the pointer pointing to the central scale value.
    BatteriesFour AA (R6P) batteries
    Approx. Width125 mm
    Approx. Height103 mm
    Approx. Depth53 mm
    Approx. Weight500 g
    ComplianceEN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)
    Standard AccessoriesBatteries (Qty - 4)
    Earth Probe (98002) (Qty - 1)
    Line Probe (98001) (Qty - 1)
    Protection Cover (93013) (Qty - 1)
    Shoulder Strap (99005) (Qty - 1)
    User's Manual (Qty - 1)


    Standard Test Conditions - Model MY10-05

    Ambient Temperature/Humidity Ranges23 ±5 /45-75% RH
    Position of UseHorizontal (5° max. of angle of inclination)
    Battery VoltageWithin effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)
    Tolerance Mentioned ConditionsResistance measurement: First effective measuring range = ±5% of reading
    Second effective measuring range = ±10% of reading
    Infinite and zero scale values: 0.7% max. of scale length
    AC voltage: ±10% of maximum scale value
    No-load voltage: 130% max. of rated voltage
    Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range
    Short-circuit current: 12 mA max.


    Additional Functions - Model MY10-05

    • Automatic discharge function
      • If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
      • When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.
    • Battery check (BAT mark on the scale plate)